Performance features:
The rapid automatic hydrogen analyzer features an integrated design with an elegant appearance, lightweight construction, and excellent maintainability. The entire testing process is controlled by a microcontroller that processes and calibrates test data. It displays the milligram concentration of hydrogen in the sample and prints the measurement results. Each sample analysis takes approximately 10-15 minutes, with accuracy meeting the requirements of China's national standard GB/T15460-2003. Additionally, the instrument can measure carbon using the gravimetric absorption method, ensuring accuracy that complies with the same standard.
※ No restriction on internal water input (internal water>10, results are directly available), effectively preventing experimental failure caused by excessive sample moisture.
Technical parameter:
Measurement range: Hydrogen 0-20%, Carbon 1-100
Rise rate: 25℃/min, 30min or 800℃
The determination time was about 10-15 minutes, including 30 seconds at 300℃,2 minutes at 500℃ and 10 minutes at 850℃.
Measurement accuracy: meets the requirements of GB/T15460—2003
Allowable error: Hydrogen: ≤0.15% (within the same laboratory); Carbon: ≤0.50%
The weight of the analyzed coal sample with particle size <0.2mm and air-dried is about 65mg-75mg.
Instrument linearity: ±0.1%
The heating rate was about 50℃/min.
Electrolysis voltage: 10V
Working electrolysis voltage: 24V
Current: Electrolysis current>600mA, final current <70mA
Power supply: AC220V±10%,50Hz
External dimensions: 1170mm × 500mm × 400mm;
Total weight: 35kg