Hydrolysis partial function features:
The instrument is equipped with a 7-inch LCD touch screen displaying real-time temperature, sample delivery position and time. The interface is user-friendly and the operation is simple.
The furnace temperature is controlled by PID flow regulation, which ensures high precision and extends the service life of silicon carbide tubes.
Fully automatic sample loading, dispensing, sealing, and plugging with automatic reset.
The high precision digital electrode was configured to ensure the accuracy of the experiment.
Service condition:
(1) temperature :0-40℃
(2) Relative humidity: ≤80%
(3) Power supply: Voltage: 220±22V
Frequency: 50±0.5Hz
(4) Carrier gas: helium, with a purity of over 99.8%.
(5) Power: No less than 4.5KW.
(6) No interference from strong magnetic fields
Basic parameters of nitrogen analyzer:
(1) The measurement range is 0.01-50%.
(2) Sample quantity: about 100mg;
(3) The single measurement time was about 35 min.
(4) Rated power consumption: 4000W
Technical requirements for nitrogen determinators:
(1) Repeatability (in accordance with GB/T 19227-2008) ≤0.08%
(2) Temperature control accuracy: ±10℃
(3) Temperature control range:
hydrolysis furnace: room temperature to 1200℃
STEAM GENERATOR: room temperature to 500℃
Distillation furnace: room temperature to 500℃